The aim of this study was to evaluate the effect of inhomogeneities on dose calculation for low energy photons intra-operative radiation therapy (IORT) in pelvic area. A GATE Monte Carlo model of the INTRABEAM® was adapted for the study. Simulations were performed in the CT scan of a cadaver considering a homogeneous segmentation (water) and an inhomogeneous segmentation (5 tissues from ICRU44). Measurements were performed in the cadaver using EBT3 Gafchromic® films. Impact of inhomogeneities on dose calculation in cadaver was 6% for soft tissues and greater than 300% for bone tissues. EBT3 measurements showed a better agreement with calculation for inhomogeneous media. However, dose discrepancy in soft tissues led to a sub-millimeter (0.65âmm) shift in the effective point dose in depth. Except for bone tissues, the effect of inhomogeneities on dose calculation for low energy photons intra-operative radiation therapy in pelvic area was not significant for the studied anatomy.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Sophie Chiavassa, François Buge, Chloé Hervé, Gregory Delpon, Jérome Rigaud, Albert Lisbona, Stéphane Supiot,