Article ID Journal Published Year Pages File Type
10731042 Physica Medica 2015 7 Pages PDF
Abstract
The aim of this study was to evaluate the effect of inhomogeneities on dose calculation for low energy photons intra-operative radiation therapy (IORT) in pelvic area. A GATE Monte Carlo model of the INTRABEAM® was adapted for the study. Simulations were performed in the CT scan of a cadaver considering a homogeneous segmentation (water) and an inhomogeneous segmentation (5 tissues from ICRU44). Measurements were performed in the cadaver using EBT3 Gafchromic® films. Impact of inhomogeneities on dose calculation in cadaver was 6% for soft tissues and greater than 300% for bone tissues. EBT3 measurements showed a better agreement with calculation for inhomogeneous media. However, dose discrepancy in soft tissues led to a sub-millimeter (0.65 mm) shift in the effective point dose in depth. Except for bone tissues, the effect of inhomogeneities on dose calculation for low energy photons intra-operative radiation therapy in pelvic area was not significant for the studied anatomy.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , ,