Article ID Journal Published Year Pages File Type
10731064 Physica Medica 2015 4 Pages PDF
Abstract
In the first stage of this study, a suitable detector was chosen and its response vs. activity was characterized. The experimental verification was performed measuring the ambient dose equivalent rate from patients receiving radioiodine administration. The results confirm the reliability of the proposed method, as shown by the slight differences between the administered activities and the ones calculated from external measurements. Furthermore, the same procedure was applied to detect the percentage residual activity in patients at two preset time intervals: 4 hours and 4 days after the radioiodine administration. The obtained results clearly highlight that the method can ensure a level of reliability compatible with the radiation protection purposes.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , ,