Article ID Journal Published Year Pages File Type
10731084 Physica Medica 2015 7 Pages PDF
Abstract
We confirmed ray-tracing in P3DS with optic simulation and found that 2D projection and 3D reconstructed images were qualitatively similar to the phantom images. For pre-clinical tests, the dose distribution and profile showed good agreement among RTP, optical CT, and external beam radiotherapy film data for the axial and coronal views. The P3DS has shown that it can scan and reconstruct for evaluation of the gel dosimeter within 1 min. We confirmed that the P3DS system is a useful tool for the measurement of 3D dose distributions for 3D radiation therapy QA. Further experiments are needed to investigate quantitative analysis for 3D dose distribution.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , , , , , , , , , , , ,