Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10731106 | Physica Medica | 2005 | 10 Pages |
Abstract
Applying a correct quality assurance program based on the tests of the patient set-up, machine settings and calculations, results showed that the method is able to verify agreements between the dose Dm,TPS and the in-vivo dose value Dm, within 4% for 95% of the 240 measurements carried out in-vivo.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Angelo Piermattei, Luca Grimaldi, Giuseppe D'Onofrio, Savino Cilla, Pietro Viola, Maurizio Craus, Andrea Fidanzio, Luigi Azario, Francesco Deodato, Gabriella Macchia, Alessio Morganti,