Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10731350 | Physica Medica | 2016 | 5 Pages |
Abstract
Record and verify (R&V) systems have proven that their application in radiotherapy clinics leads to a significant reduction in mis-treatments of patients. The purpose of this technical note is to share our experience of acceptance testing, commissioning and setting up a quality assurance programme for the MOSAIQ® oncology information system and R&V system after upgrading from software version 2.41 to 2.6 in a multi-vendor, multi-site environment. Testing was guided primarily by the IAEA Human Report No. 7 recommendations, but complemented by other departmental workflow specific tests. To the best of our knowledge, this is the first time successful implementation of the IAEA Human Health Report Series No. 7 recommendations have been reported in the literature.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Thulani Nyathi, Christopher Colyer, Anup Kumar Bhardwaj, James Rijken, Jason Morton,