Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10731478 | Physica Medica | 2014 | 4 Pages |
Abstract
Independent monitor unit verification calculation (MUVC) has been recommended by several authors for intensity modulated radiotherapy (IMRT) as a patient specific quality assurance tool. Aim of the present work is to develop an in-house excel spread sheet based MUVC program for volumetric modulated arc therapy (VMAT) using Clarkson's integration technique. Total scatter factor (Sc,p) and tissue maximum ratio (TMR) for circular fields obtained from Treatment planning system (TPS) were used for the calculation. Multileaf collimator (MLC) interleaf leakage, MLC round edge transmission and tongue and groove effect were accounted. MUVC calculation was performed for 58 patients both for patient anatomy and for homogenous cylindrical phantom. Radiological path lengths were used as water equivalent depths (WED) for calculations using patient anatomy. Monitor unit (MU) discrepancies between â2.60% and 0.28% with mean deviation of â0.92% ± 0.75% were obtained for homogenous cylindrical phantom calculations. MUVC for patient anatomy resulted in large variations between â19.02% and 0.67% for 14 plans where isocenter was at a region below â350 HU. But For 44 plans where the isocenter was at a region above â350 HU, variations between â3.44% and 0.48% were obtained with mean deviation of â1.73% ± 1.12%. For VMAT patient specific quality assurance, the independent MUVC algorithm can be used as an easy and quick auxiliary to measurement based verification for plans with isocenter at a region above â350 HU.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Prakash Jeevanandam, Dhanabalan Rajasekaran, Prabakar Sukumar, Vivekanandan Nagarajan,