Article ID Journal Published Year Pages File Type
10731727 Radiation Measurements 2005 7 Pages PDF
Abstract
A mathematical model, based on Monte Carlo simulation, is proposed for deriving absorbed energy and dose distribution in mammography utilizing a mathematical water-like phantom. The model was validated for its accuracy against experimental and published data. The main factor discriminating absorbed energy distribution characteristics among different mammographic techniques was considered the X-ray spectrum. The absorbed energy distribution inside the phantom was investigated via percentage depth dose and isodose curves. The influence of the factors affecting X-ray spectrum (tube voltage, anode material, filter material and thickness) on absorbed energy distribution was examined. The hardness of the beam, due to increase of tube voltage or filtration, was found to be the major factor affecting absorbed energy distribution inside the phantom. In general, Mo and W anode systems demonstrated superior dosimetric characteristics against those of W-Mo or Rh. The model presented can be used for estimating absolute and relative breast dose values and their spatial distributions.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , ,