Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10916177 | Nuclear Medicine and Biology | 2005 | 6 Pages |
Abstract
A conventional high-performance liquid chromatographic (HPLC) method for the analysis of 2-fluoro-2-deoxy-d-glucose (FDG) and 2-deoxy-2-chloro-d-glucose (ClDG) in [18F]FDG preparations is described. This method was based on a postcolumn derivatization with 2-cyanoacetamide (2-CA) and UV detection. FDG and ClDG were separated on a normal-phase column using acetonitrile/water as the mobile phase. The eluate was mixed with 2-CA in sodium borate buffer solution at the outlet of a PTFE coil (10 mÃ0.5 mm id) from the column, and the reaction was carried out at 100°C during the passage through the coil. The UV absorbance of the resultant product was monitored at 276 nm. Under optimum conditions, the detection limits [signal-to-noise (S/N) ratio=3] for FDG and ClDG were 0.31 and 0.17 μg/ml for a 20-μl injection volume, respectively, and the linearity ranges were 0.5-100 μg/ml for both compounds. The intra- and interday reproducibilities were better than 2.2% [relative standard deviation (R.S.D.)]. This HPLC separation procedure is also useful for determining the radiochemical purity of [18F]FDG preparations since it allows the analysis of 2-[18F]fluoro-1,3,4,6-tetra-O-acetyl-d-glucose ([18F]TAG), partially hydrolyzed [18F]TAG and [18F]Fâ. This method can be used at many positron emission tomography (PET) facilities since it does not require an expensive, sophisticated electrochemical detector.
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Cancer Research
Authors
Ryuji Nakao, Takehito Ito, Masatoshi Yamaguchi, Kazutoshi Suzuki,