Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
10918233 | Radiotherapy and Oncology | 2015 | 6 Pages |
Abstract
Dual-quadratic CyberKnife and 2D Vero correlation models were interchangeable in terms of geometrical accuracy with the CK linear ME95,BEVÂ =Â 4.1Â mm, CK dual-quadratic ME95,BEVÂ =Â 3.9Â mm and Vero ME95,BEVÂ =Â 3.7Â mm, when modeled with FF sequence. CK DQ modeling based on 15p acquired in 20 s may lead to problems for internal motion estimation.
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Cancer Research
Authors
Kenneth Poels, Jennifer Dhont, Dirk Verellen, Oliver Blanck, Floris Ernst, Jef Vandemeulebroucke, Tom Depuydt, Guy Storme, Mark De Ridder,