Abstract
												Intrafraction change in tumor position (Î) was evaluated for stereotactic lung radiotherapy delivered with flattening filter free volumetric modulated arc therapy. In 140 fractions from 32 patients mean Î (±SD) was â0.7 ± 1.4 mm (vertical), â0.7 ± 1.3 mm (longitudinal) and +0.2 ± 1.2 mm (lateral) with mean vector 2.1 ± 1.2 mm. Mean delivery time was 4.4 ± 3.4 min (mean beam-on 1.9 ± 0.4 min).