Article ID Journal Published Year Pages File Type
1162743 Analytica Chimica Acta 2016 6 Pages PDF
Abstract

•Fast, accurate, and precise water soluble vitamin determination in a wide range of infant formula and nutritional products.•Improvements in sample preparation and separation conditions are key contributors to method performance.•The method was validated in eleven different infant formula and related matrixes with excellent demonstrated accuracy and precision.

An improved method was developed for simultaneous determination of the fortified forms of thiamine (B1), riboflavin (B2), nicotinamide and nicotinic acid (B3), pantothenic acid (B5), pyridoxine (B6), biotin (B7), and folic acid (B9) in infant formulas and related nutritionals. The method employed a simple, effective, and rapid sample preparation followed by liquid chromatography tandem mass spectrometry (LC-MS/MS). It improved upon previous methodologies by offering facile and rugged sample preparation with improved chromatographic conditions, which culminated in a highly accurate and precise method for water-soluble vitamin determination in a wide range of formulas. The method was validated over six days in ten unique matrices with two analysts and on instruments in two different labs. Intermediate precision averaged 3.4 ± 2.6% relative standard deviation and over-spike recovery averaged 100.2 ± 2.4% (n = 160). Due to refinements in sample preparation, the method had high sample throughput capacity.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,