Article ID Journal Published Year Pages File Type
1174244 Analytical Biochemistry 2011 7 Pages PDF
Abstract
Neutral filter elution assay is one of the methods used for detection of DNA double strand breaks (DSBs). However, it is laborious, expensive, and hazardous (radiolabeled precursors for DSB detection and scintillation counter for quantification), making it a less preferred method for DSB detection. In the present study, an attempt was made to improve the existing neutral filter elution assay by making use of fluorescent dye (PicoGreen) and microfiltration assembly for eluting the fragmented DNA, thereby reducing the cost and time required for the assay. We studied the effect of dye dilution, pH conditions, and cell number as a part of method standardization. X-ray dose-response and repair kinetics in lymphocytes as well as cell lines were studied for validating the sensitivity of the assay. A linear dose-response relationship for DSBs was observed at a cell number of 4 × 105 cells, a dye dilution of 500-fold, and at pH 10. Repair kinetics revealed a time-dependent repair of DSBs up to 360 min of posttreatment, indicating its usefulness in DSB repair studies. In conclusion, the present modified method is more efficient (in terms of cell number), cost effective, less time-consuming, and less hazardous compared to the existing method.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,