Article ID Journal Published Year Pages File Type
1175319 Analytical Biochemistry 2009 8 Pages PDF
Abstract

We report a three-phase chromatographic method for the separation and analysis of δ13C values of underivatized amino acids from biological proteins (keratin, collagen, and casein) using liquid chromatography–isotope ratio mass spectrometry (LC–IRMS). Both precision and accuracy of δ13C values for standard amino acid mixtures over the range of approximately 8 to 1320 ng of carbon per amino acid on the column were assessed. The precision of δ13C values of amino acids was found to be better at higher concentrations, whereas accuracy improved at lower concentrations. The optimal performance for this method was achieved with between 80 and 660 ng of carbon of each amino acid on the column. At amino acid amounts lower than 20 ng of carbon on the column, precision and accuracy may become compromised. The application of this new three-phase chromatographic technique will allow the analysis of δ13C of amino acids to be carried out as a routine method and benefit fields of research such as biomedicine, forensics, ecology, nutrition, and palaeodiet reconstruction in archaeology.