Article ID Journal Published Year Pages File Type
1177329 Analytical Biochemistry 2007 7 Pages PDF
Abstract

Based on the isopeptidase activity of human plasma FXIII, a novel fluorometric assay that determines FXIII concentrations in human plasma below 0.05 IU/ml is introduced. We considered a peptide sequence derived from α2-antiplasmin (n = 12) to yield high sensitivity. Peptide Abz-NE(Cad-Dnp)EQVSPLTLLK exhibits a Km value of 19.8 ± 2.8 μM and is used in a concentration of 50 μM. The assay design is suitable for measurements in cuvettes (1 ml volume) as well as for the microtiter plate (MTP) format (0.2 ml volume). It provides linear dose–response curves over a wide range of FXIII concentrations (0.05–8.8 IU/ml). The assay was validated with respect to precision, detection and quantitation limits, accuracy/specificity, linearity, and range. A comparison of the fluorometric assay with the photometric assay for FXIII determinations in plasma pools as well as single donor plasma revealed suitability of the fluorometric assay for FXIII determination in plasma of healthy individuals. FXIII concentrations in plasma samples of patients with severe FXIII deficiency are discussed in the context of FXIII antigen levels. These assays correlate well in the critical range below 0.1 IU/ml, whereas the photometric assay may overestimate residual FXIII activity in severe FXIII-deficient patients.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,