Article ID Journal Published Year Pages File Type
1183654 Chinese Journal of Chromatography 2008 5 Pages PDF
Abstract

A method for the determination of ifenprodil levels in human plasma was established. Ifenprodil and the internal standard (IS), ketoconazole, were extracted from the plasma with ethyl acetate using liquid-liquid extraction. The extracts were separated by high performance liquid chromatography (HPLC), using the mobile phase of methanol-6 mmol/L ammonium acetate (90:10, v/v) (pH value was adjusted to 7.40). The electrospray ion source was applied and operated in a positive ion mode under the selected reaction monitoring mode with a transition of m/z 326.1→308.2, for quantification of ifenprodil, and m/z 531.0→82.1 for IS. The liquid chromatography-tandem mass spectrometry method provided excellent sensitivity and selectivity for quantification and identification of ifenprodil at low levels, within a run time of 6.0 min. The assay was linear over the range from 0.25 to 50 μg/L. The relative standard deviations for the intra-day and inter-day precision were less than 2.7% and 6.5%, respectively. The average recoveries varied between 101.3% and 105.0%, and the detection limit was 0.08 μg/L. Because of its simplicity and accuracy, the established method is suitable for clinical application in the determination of ifenprodil in human plasma.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry