Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1193295 | International Journal of Mass Spectrometry | 2009 | 6 Pages |
Abstract
A pulsed ion source has been developed supplying two slightly shifted mass spectra for every single ionization pulse, e.g., by laser induced resonance enhanced multiphoton ionization (REMPI). This allows comparison of ion signals with a difference of relative intensity in the percent and sub-percent range. Such small differences may be due to small changes of gas composition or of ionization conditions and usually are hidden by pulse to pulse fluctuations of laser pulse energy, gas density, electric fields, ion detection, etc. With the presented twin-peak ion source, both ion signals are subject to the same fluctuations which will influence the ratio of the ion signals only marginally. This method has been applied to circular dichroism laser mass spectrometry, where relative ion signal ratios 2(Ir â Il)/(Ir + Il) due to ionization with right and left circularly polarized light have to be recorded in the 10â3 range and below.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Christoph Logé, Alexander Bornschlegl, Ulrich Boesl,