Article ID Journal Published Year Pages File Type
1195671 Journal of the American Society for Mass Spectrometry 2006 8 Pages PDF
Abstract

The number of possible chemical formulae assigned to an accurate determined mass was significantly reduced by comparing spectral and theoretical isotope patterns based on mass measurement obtained with an ultrahigh-resolution electrospray ionization Fourier transform ion cyclotron resonance mass spectrometer (ESI-FTICR-MS) at high field intensity (7 T). Reduction is performed by rating congruency between experimental and theoretical pattern intensity and mass, and filtering out compositions with insufficient user-definable results. The methods used for isotope pattern simulation, peak searching, and comparison will be briefly described and evaluated on molecule ion signals of 25 compounds (300–1000 Da) applying a mass accuracy of ±5 ppm, a set of eight elements with constant constraints (C0-200H0-1000N0-15O0-15S0-2Cl0-2Br0-2Ru0-1), natural isotope abundances and experimental resolution (full width at half maximum).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,