Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1196047 | Journal of the American Society for Mass Spectrometry | 2008 | 10 Pages |
A novel gas chromatograph-mass spectrometer (GC-MS) based on a miniature toroidal ion trap mass analyzer (TMS) and a low thermal mass GC is described. The TMS system has an effective mass/charge (m/z) range of 50–442 with mass resolution at full-width half-maximum (FWHM) of 0.55 at m/z 91 and 0.80 at m/z 222. A solid-phase microextraction (SPME) fiber mounted in a simple syringe-style holder is used for sample collection and introduction into a specially designed low thermal mass GC injection port. This portable GC-TMS system weighs <13 kg (28 lb), including batteries and helium carrier gas cartridge, and is totally self-contained within dimensions of 47 × 36 × 18 cm (18.5 × 14 × 7 in.). System start-up takes about 3 min and sample analysis with library matching typically takes about 5 min, including time for column cool-down. Peak power consumption during sample analysis is about 80 W. Battery power and helium supply cartridges allow 50 and 100 consecutive analyses, respectively. Both can be easily replaced. An on-board library of target analytes is used to provide detection and identification of chemical compounds based on their characteristic retention times and mass spectra. The GC-TMS can detect 200 pg of methyl salicylate on-column. n-Butylbenzene and naphthalene can be detected at a concentration of 100 ppt in water from solid-phase microextraction (SPME) analysis of the headspace. The GC-TMS system has been designed to easily make measurements in a variety of complex and harsh environments.
Graphical AbstractThe design and performance of a self-contained, hand-portable gas chromatograph-toroidal ion trap mass spectrometer for field measurements are described.Figure optionsDownload full-size imageDownload high-quality image (190 K)Download as PowerPoint slide