Article ID Journal Published Year Pages File Type
1199538 Journal of Chromatography A 2015 11 Pages PDF
Abstract

•Comparison the cleanup performance of d-SPE and m-PFC on pesticide multiresidue analysis.•Improved previous m-PFC columns to simplify the equipment.•M-PFC had more stable recoveries and lower RSDs.•M-PFC method reduced matrix interferences and had less matrix effects.•M-PFC is a rapid and high-throughput cleanup method on analysis of pesticides.

In this study, dispersive-Solid Phase Extraction (d-SPE) cleanup and multi-plug filtration cleanup (m-PFC) methods were compared for 25 representative pesticides in six matrices (wheat, spinach, carrot, apple, citrus and peanut) by QuEChERS-LC-ESI-MS/MS detection. The type of sorbents in dispersive-SPE (d-SPE) was optimized for the above matrices. Multi-walled carbon nanotubes (MWCNTs), which mixed other materials like PSA (Primary Secondary Amines), GCB (Graphitized Carbon Black) and C18 (Octadecyl-silica), showed brilliant cleanup performance in multi residue monitoring (MRM) pesticide residue analysis. Cleanup effects with d-SPE and m-PFC methods were examined. When spiked at 3 concentration levels of 10, 100, 500 μg/kg in above matrices, for both d-SPE and m-PFC methods, the recoveries ranged from 70 to 110% with relative standard deviations (RSDs) lower than 20%. Limits of quantification (LOQs) for both cleanup methods ranged from 1 to 25 μg/kg. Matrix-matched calibrations were performed with the coefficients of determination more than 0.99 between concentration levels of 10–1000 μg/kg. It was found that m-PFC was more convenient and effective than d-SPE with the same sorbents, due to the increased contact time and contact area between the extracts and compressed sorbents. The study demonstrated that m-PFC method could be used as a rapid, convenient and high-throughput cleanup method for analysis of pesticide residues.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , , , , ,