Article ID Journal Published Year Pages File Type
1199895 Journal of Chromatography A 2014 9 Pages PDF
Abstract

•New hybrid dummy molecularly imprinted polymers.•3-Aminopropyltriethoxy silane-methacrylic acid as hybrid monomer.•Miniaturized molecularly imprinted solid-phase extraction in 1.5-mL tapered centrifuge tube.•Good recognition and selective extraction of acyclovir from urine.

The miniaturized molecularly imprinted solid-phase extraction (mini-MISPE) coupled with high-performance liquid chromatography was proposed for the determination of acyclovir in urine. 1.5-mL tapered plastic centrifuge tube filled with hybrid molecularly imprinted polymers (HMIPs) was used as the cartridge of mini-MISPE, and the HMIPs synthesized with 3-aminopropyltriethoxy silane-methacrylic acid as monomer exhibited good recognition and selectivity for acyclovir. Under the optimized condition, good linear calibration was obtained in a range of 0.5–15 μg mL−1 with the correlation coefficient of 0.9994, and the recoveries at three spiked levels were 91.6–103.3% in urine with the relative standard deviation (RSD) of ≤3.5%. Excellent intra-day and inter-day repeatability were achieved with RSD of ≤2.6% and 4.0% in three different concentrations. This method combined the advantages of HMIPs and mini-MISPE, and it could become an alternative tool for analyzing the residues of acyclovir in complex urine matrices.

Keywords
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,