Abstract
N-acetylneuraminic acid (Neu5Ac) and N-acetylglycolylneuraminic acid (Neu5Gc), two acylated derivatives of 9-C carboxylated monosaccharides, are involved in a number of biological processes as modulators of glycoconjugates. A partially automated method is here presented for determination of these sialic acids in the two most important biofluids for clinical analysis: serum and urine. For this purpose, a solid-phase extraction (SPE) workstation was on-line connected to an LC-MS/MS triple quadrupole mass detector. Hydrolysis to release sialic acids bound to glycoconjugates and derivatization were the two steps implemented as sample preparation prior to SPE-LC-MS/MS analysis. Following thorough optimization of the SPE and LC-MS/MS conditions, the analytical method was validated using the standard addition approach to assess the presence of matrix effects. The proposed method affords detection limits of 0.03Â ng/mL and 0.04Â ng/mL for Neu5Ac and Neu5Gc, respectively. The precision (expressed as relative standard deviation) was 1.7 and 4.6% for within-day variability, and 4.8 and 7.2% for between-days variability. Accuracy, estimated using spiked (between 1 and 50Â ng/mL) and non-spiked samples of both biofluids, ranged from 95.2 to 99.6%. The method was applied to human serum and urine of healthy volunteers, thus showing its suitability for application in both clinical and research laboratories.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
F. Priego-Capote, M.I. Orozco-Solano, M. Calderón-Santiago, M.D. Luque de Castro,