Article ID Journal Published Year Pages File Type
1200490 Journal of Chromatography A 2016 8 Pages PDF
Abstract

•The analysis avoids derivatization and complex sample preparation steps.•Only 150 μL of samples were needed for various types of water analysis.•Both primary and secondary quantitative limits of glyphosate and AMPA are 0.1 ppb.•This method meets requirements for EU standards.

A novel method based on high performance liquid chromatography coupled to tandem mass spectrometry (HPLC-MS/MS) was developed for the sensitive determination of glyphosate and its major degradation product, AMPA in environmental water samples. The method involves the use of MS compatible mobile phases (0.1% formic acid in water and acetonitrile) for HPLC and direct analysis of water samples without sample derivatization. The method has been validated in different types of water matrices (drinking, surface and groundwater) by accuracy and precision studies with samples spiked at 0.1, 7.5 and 90 ppb. All mean accuracy values ranged from 85% to 112% for glyphosate and AMPA using both primary and secondary quantitative ion transitions (RSD ≤ 10%). Moreover, both primary and secondary ion transitions for glyphosate and AMPA can achieve the quantitation limits at 0.1 ppb. The linear dynamic range of the calibration curves were from 0.1 to 100 ppb for each analyte at each ion transitions with correlation coefficient higher than 0.997.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,