Article ID Journal Published Year Pages File Type
1201069 Journal of Chromatography A 2015 6 Pages PDF
Abstract
Portable and compact instruments for separating and detecting organic compounds are needed in the field for environmental studies. This is especially the case for pollution studies as in-field detection of organic compounds helps identify sources of pollution. Here we describe the development of a compact GC and simple pre-concentrator coupled to a MS detector. This simple system can easily be incorporated into portable instrumentation. Combining the pre-concentrator and compact column has the advantage of decoupling separation from manual injection and enhances separation of environmentally relevant polar organic compounds, such as levoglucosan. A detection limit of 2.2 ng was obtained for levoglucosan. This simple design has the potential to expand the use of gas chromatography as a routine in-field separation technique.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,