Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1203020 | Journal of Chromatography A | 2010 | 4 Pages |
The temperature effects during the sol–gel process and ageing of the silica-based monolith on the structure and separation efficiency of the capillary columns (100 μm i.d., 150 mm) for HPLC separations were studied. The tested columns were synthesized from a mixture of tetramethoxysilane, polyethylene glycol and urea under the acidic conditions. The temperature was varied from 40 °C to 44 °C and formation of bypass channels between the silica mold and the capillary wall was examined. The temperature of 43 °C was estimated as optimal for preparation of efficient silica capillary columns which were subsequently modified by octadecyldimethyl-N,N-diethylaminosilane or covered by poly(octadecyl methacrylate) and tested using standard mixture of alkylbenzenes under the isocratic conditions.