Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1203248 | Journal of Chromatography A | 2010 | 9 Pages |
Two-dimensional gas chromatography (GC × GC) coupled with time-of-flight mass spectrometric (TOFMS) method was optimized for simultaneous analysis of 160 pesticides, 12 dioxin-like polychlorinated biphenyls (PCBs), 12 polyaromatic hydrocarbons (PAHs) and bisphenol A in grape and wine. GC × GC–TOFMS could separate all the 185 analytes within 38 min with >85% NIST library-based mass spectral confirmations. The matrix effect quantified as the ratio of the slope of matrix-matched to solvent calibrations was within 0.5–1.5 for most analytes. LOQ of most of the analytes was ≤10 μg/L with nine exceptions having LOQs of 12.5–25 μg/L. Recoveries ranged between 70 and 120% with <20% expanded uncertainties for 151 and 148 compounds in grape and wine, respectively, with intra-laboratory Horwitz ratio <0.2 for all analytes. The method was evaluated in the incurred grape samples where residues of cypermethrin, permethrin, chlorpyriphos, metalaxyl and etophenprox were detected at below MRL.