Article ID Journal Published Year Pages File Type
1203350 Journal of Chromatography A 2010 14 Pages PDF
Abstract

A hydrophilic interaction liquid chromatography (HILIC) method was used to separate a commonly used pharmaceutical starting material, 4-aminomethylpyridine (4-AMP), and its degradants. The structures of the major degradants were characterized and elucidated without prior isolation by accurate mass measurement, MS/MS analysis and on-line hydrogen/deuterium (H/D) exchange experiments. The mass spectra obtained from H/D exchange experiments are particularly useful to differentiate structural isomers, to elucidate the fragmentation pathways, and to aid in structure elucidation in the absence of MS/MS fragmentation information. The impact of deuterium oxide and temperature on HILIC separation has also been explored here. The integration of H/D exchange with HILIC has been described here for the first time and has been demonstrated to be a powerful structure elucidation tool via the study of degradants in 4-AMP.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,