Article ID Journal Published Year Pages File Type
1204744 Journal of Chromatography A 2009 7 Pages PDF
Abstract

Pressure–flow curves are obtained for a new protein A adsorbent matrix based on macroporous hydrophilic polymer beads with average diameter of 57 μm and a narrow particle size distribution. Experimental data are obtained in a 1 cm diameter laboratory column and in preparative scale columns with diameters of 20, 30, and 45 cm. The results are consistent with a model that assumes a linear relationship between bed compression and relative flow velocity. Surprisingly, the packing compressibility is essentially independent of column diameter for the preparative columns. As a result, after accounting for the variation in extraparticle porosity caused by compression, the column pressure drop is accurately predictable using the Carman–Kozeny equation. A model is also developed to predict productivity for IgG capture as a function of operating conditions based on dynamic binding capacity data presented in Part I of this work. For typical conditions, the model predicts maximum productivity at low residence times, between 1 and 1.5 min, when the dynamic binding capacity is at about 70–80% of the maximum. Combining the two models for column pressure and for dynamic binding capacity allows the design of preparative scale columns that maximize productivity while meeting specified pressure constraints.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,