Article ID Journal Published Year Pages File Type
1204940 Journal of Chromatography A 2010 8 Pages PDF
Abstract

A rapid, precise and fully-automated method for analysis of folate (vitamin B9) and its catabolites (viz. p-aminobenzoylglutamate and its acetamide derivative) in biofluids is here presented. The method is based on on-line hyphenation of solid-phase extraction (SPE) by a Prospekt 2 system with hydrophilic interaction liquid chromatography–tandem mass spectrometry (HILIC–MS/MS). The method was analytically characterized by estimation of repeatability (RSD, n = 5, between 0.5 and 4.1%), accuracy (between 96 and 105%), and sensitivity (limits of quantificantion between 0.3 and 8.3 ng/mL (1.1 and 18.8 pmol/mL) or 0.03 and 0.83 ng (0.11 and 1.88 pmol)). The proposed method is suited for routine analysis of folate catabolites as biomarkers to monitor deficiency of vitamin B9.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,