Article ID Journal Published Year Pages File Type
1205956 Journal of Chromatography A 2008 5 Pages PDF
Abstract

In this work, a piece of glass fiber was inserted into the channel of a poly(methyl methacrylate) (PMMA) electrophoresis microchip to enhance the electroosmotic flow (EOF) and the separation efficiency. The EOF value of the glass fiber-containing microchannel at pH 8.2 was determined to be 4.17 × 10−4 cm2 V−1 s−1. The performance of the new microchip was demonstrated by its ability to separate and detect three purines coupled with end-column amperometric detection. In addition, a piece of trypsin-immobilized glass fiber was inserted into the channel of a PMMA microchip to fabricate a core-changeable microfluidic bioreactor that can be regenerated by changing the fiber. The in-channel fiber bioreactor has been coupled with matrix-assisted laser desorption ionization time-of-flight mass spectrometry for the digestion and peptide mapping of bovine serum albumin and myoglobin.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,