1206961 | Journal of Chromatography A | 2009 | 7 Pages |
A new method for the measurement of N-nitrosamines in part-per-trillion concentrations from water samples without preconcentration steps has been developed. This method is based on online UV irradiation after high-performance liquid chromatographic separation and subsequent luminol chemiluminescence detection without addition of an oxidant. It was confirmed that N-nitrosamines in basic aqueous solution were transformed to peroxynitrite by UV irradiation. The detection limits for this method were 1.5 ng/L, 2.9 ng/L, 3.0 ng/L, and 2.7 ng/L for N-nitrosodimethylamine, N-nitrosomorpholine, N-nitrosomethylethylamine, and N-nitrosopyrrolidine, respectively, at a signal-to-noise ratio of 3. The calibration graphs were linear in the range of 5–1000 ng/L for these N-nitrosamines. This method was used for the determination of N-nitrosamines in tap water, river water, and industrial plant effluent samples. The recoveries of N-nitrosodimethylamine, N-nitrosomorpholine, N-nitrosomethylethylamine, and N-nitrosopyrrolidine present in tap water sample at a concentration of 10 ng/L (mean ± standard deviation, n = 4) were (94.8 ± 2.7)%, (102.0 ± 6.9)%, (99.3 ± 3.9)%, and (102.8 ± 2.5)%, respectively. These results indicate that our proposed method can be applied satisfactorily to the determination of N-nitrosamines in water samples.