Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1207337 | Journal of Chromatography A | 2007 | 6 Pages |
Abstract
A new non competitive capillary electrophoresis immunoassay format based on a separation into a capillary modified by analyte immobilisation is described. The injection of an excess of labelled antibody off-line preincubated with the analyte allows the surface capture of the free antibody and the immunocomplex detection. It was developed using the human serum albumin (HSA) as analyte, two FITC-labelled antibodies and a HSA covalently linked capillary. Two calibration curves with good run-to-run reproducibility and LOD – respectively 14.0 nM for the FITC-polyclonal antibody and 9.0 nM for the FITC-monoclonal antibody – were achieved. The assay was applied to HSA determination in spiked samples of human urine with acceptable recoveries.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
C. Giovannoli, L. Anfossi, C. Baggiani, G. Giraudi,