Article ID Journal Published Year Pages File Type
1207364 Journal of Chromatography A 2008 4 Pages PDF
Abstract

A simple and highly sensitive method called stir bar sorptive extraction (SBSE) and thermal desorption (TD)-gas chromatography–mass spectrometry (GC–MS), which is used for the determination of trace amounts of 5-chloro-2-(2,4-dichlorophenoxy)phenol (triclosan) in river water samples, is described. A stir bar coated with polydimethylsiloxane (PDMS) is added to a 10 mL water sample and stirring is carried out for 120 min at room temperature (25 °C) in a vial. Then, the PDMS stir bar is subjected to TD–GC–MS. The detection limit of triclosan is 5 ng L−1 (ppt). The method shows linearity over the calibration range (0.02–20 μg L−1) and the correlation coefficient is higher than 0.997 for triclosan standard solution. The recovery of triclosan in river water samples ranges from 91.9 to 108.3% (RSD: 4.0–7.0%). This simple, accurate, sensitive, and selective analytical method may be used in the determination of trace amounts of triclosan in river water samples.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , ,