Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1207373 | Journal of Chromatography A | 2009 | 8 Pages |
A new approach using single-drop microextraction (SDME) and gas chromatography–mass spectrometry for the determination of six haloacetic acids (HAAs) in water samples was presented. n-Octanol was used as extractant and derivatization reagent. HAAs were derivatized both simultaneously during the extraction in the solvent microdrop, and after extraction, inside a glass microvial (1.1 mm I.D.). Trifluoroacetic anhydride (TFAA) was used as the reaction catalyst. The influence of catalyst amount, derivatization time and temperature on the yield of the in-microvial derivatization was investigated. Derivatization reaction was performed using 1.2 μL of TFAA at 100 °C for 20 min. Extraction was performed using 1.8 μL of n-octanol containing TFAA (10%, v/v). Experimental parameters, such as, exposure time, sample pH and extraction temperature were controlled and optimized. Analytical parameters such as linearity, precision and limit of detection were also evaluated. The proposed method was proved to be a suitable analytical procedure for HAAs in water with limits of detection 0.1–1.2 μg/L. The relative recoveries range from 82.5 to 97.6% for all the target analytes. Precision values were from 5.1 to 8.5% (as intra-day relative standard deviation, RSD) and 8.8–12.3% (as inter-day RSD).