1208819 Journal of Chromatography A 2007 7 Pages PDF
Abstract

We have developed an improved analytical method for the determination of sugar phosphates using sodium carbonate (Na2CO3) for high-performance anion-exchange chromatography–pulsed amperometric detection. The target analytes were separated completely within 10 min using eluent containing 20 mM NaOH and 35 mM Na2CO3. The limit of detection (S/N = 3) and quantitation (S/N = 10) for analytes were 10–30 ng/mL and 35–100 ng/mL, respectively. Linear dynamic range was 1–30 μg/mL (r2 ≥ 0.9998). The RSDs for intra- and inter-day assays were found to be of satisfactory results (0.23–3.09%), and the recoveries from blood spots were 97.62–99.69%.