Article ID Journal Published Year Pages File Type
1208938 Journal of Chromatography A 2007 6 Pages PDF
Abstract

An analytical separation was performed on an analytical J-type counter-current chromatography (CCC) instrument using a 5.4 ml column, with a 1 ml/min mobile phase flow rate. This separation had a resolution of 0.69 and was achieved in 10 min. The same separation was performed using two 2300 ml columns connected in series at a flow rate of 850 ml/min using a production scale J-type centrifuge. This production scale separation was also obtained in 10 min with a resolution of 0.71. This represents an 850 times increase in productivity. This paper presents these separations and the underlying scale up theory.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,