Article ID Journal Published Year Pages File Type
1209807 Journal of Chromatography A 2006 6 Pages PDF
Abstract

We have developed a high-throughput, two-dimensional-mapping (isoelectric point [pI], mass-to-charge ratio [m/z]) method by combining a capillary isoelectric focusing chip sealed with removable resin tape and a matrix-assisted laser desorption/ionization time-of-flight mass spectrometer. Sample proteins are separated in a meandering channel on the chip and immediately frozen. The tape is then removed and the proteins are freeze-dried. The freeze-drying maintains the separation state of the proteins and prevents movement of the sample solution, which can reduce pI resolution. A matrix solution is then applied and mass spectrometry is carried out by laser irradiation. The whole process takes less than 70 min, more than 10 times faster than with two-dimensional, polyacrylamide gel electrophoresis.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry