Article ID Journal Published Year Pages File Type
1211132 Journal of Chromatography A 2007 7 Pages PDF
Abstract

We applied the model described in our previous paper to the rapid scale-up in the ion exchange chromatography of proteins, in which linear flow velocity, column length and gradient slope were changed. We carried out linear gradient elution experiments, and obtained data for the peak salt concentration and peak width. From these data, the plate height (HETP) was calculated as a function of the mobile phase velocity and iso-resolution curve (the separation time and elution volume relationship for the same resolution) was calculated. The scale-up chromatography conditions were determined by the iso-resolution curve. The scale-up of the linear gradient elution from 5 to 100 mL and 2.5 L column sizes was performed both by the separation of β-lactoglobulin A and β-lactoglobulin B with anion-exchange chromatography and by the purification of a recombinant protein with cation-exchange chromatography. Resolution, recovery and purity were examined in order to verify the proposed method.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,