Article ID Journal Published Year Pages File Type
1212303 Journal of Chromatography B 2015 10 Pages PDF
Abstract

•Mixed mode chromatography media was shown that different interactions are involved.•Two mixed mode were compared using different conditions.•Interactions involved were studied.•Separation efficiency was highly dependent of the media, conditions and proteins.

Mixed mode or multimodal chromatography has been developed for rational use of multiple interactions in a controlled manner, in contrast to non-specific interactions. Indeed, as the term “mixed mode” suggests, these resins allow different types of interactions within a single chromatographic medium. In this paper, HEA HyperCel™, PPA HyperCel™ mixed-mode chromatographic media have been studied. These mixed-mode sorbents typically involve hydrophobic pseudo-affinity interactions for binding and essentially ionic interactions (charge repulsion) for elution. We identified and characterized these different interactions in chromatographic experiments by exploiting specific properties of proteins using protein standards and complex mixtures. We highlighted the major intervention of at least two types of interactions in these media: hydrophobic and electrostatic interactions. We observed the behaviour of these resins at different pH, ionic strength, with different salts and buffers types and in the presence of different organic compounds.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,