Article ID Journal Published Year Pages File Type
1213497 Journal of Chromatography B 2011 10 Pages PDF
Abstract

A liquid-chromatography–tandem-mass-spectrometry method using pneumatically assisted electrospray ionisation (LC–ESI-MS/MS) was developed for the simultaneous determination of cathinone, methcathinone, ethcathinone, amfepramone, mephedrone, flephedrone, methedrone, methylone, butylone, cathine, norephedrine, ephedrine, pseudoephedrine, methylephedrine and methylpseudoephedrine in human live and post-mortem whole blood. The blood proteins were precipitated by the addition of methanol, and the extract was purified by ultrafiltration. The separation of diastereomeric ephedrines was achieved on an ethyl-linked phenyl column. Matrix-matched calibrants combined with the isotope dilution of selected substances were used for quantitative analysis. The relative intra-laboratory reproducibility standard deviations were generally better than 7% at concentrations of 20 μg/L, and the mean true recoveries were 87–106% in the concentration range of 10–250 μg/L. The detection limits were in the range of 0.5–3 μg/L. The cathinones were unstable in whole blood and sample extracts under neutral conditions, but the stability could be improved by the acidification of the sample matrix.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
,