Article ID Journal Published Year Pages File Type
1213778 Journal of Chromatography B 2010 7 Pages PDF
Abstract

A high throughput analytical method using a column switching high-performance liquid chromatography combined with isotope dilution tandem mass spectrometry (column switching-HPLC–MS/MS) was developed to simultaneously quantitate the concentrations of 7 perfluoroalkyl acids (PFAAs) in serum and 3 PFAAs in breast milk samples. The sample preparation includes addition of the isotope-labelled internal standard solution to breast milk and serum, enzymatic hydrolysis and filtration of milk samples, precipitation of proteins and analysis by column switching-HPLC–MS/MS. The limits of quantitation ranged from 0.1 to 0.4 μg/l for serum and 0.02 to 0.15 μg/l for breast milk samples. The method accuracies ranged between 73.2% and 100.2% for the different analytes at two concentrations in PFAAs spiked samples. The validity of the method was confirmed by analysing 20 serum and 20 breast milk samples.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry