Article ID Journal Published Year Pages File Type
1213896 Journal of Chromatography B 2010 6 Pages PDF
Abstract

A novel, sensitive and reliable liquid chromatography–tandem mass spectrometric (LC–MS/MS) method was developed and validated for the determination of belinostat (PXD101) in human plasma. Oxamflatin was used as the internal standard. Liquid–liquid extraction of the plasma sample was performed using tert-butyl methyl ether as the organic solvent. Chromatographic separation was achieved on a BDS Hypersil C18 column (2.1 mm × 100 mm, 5 μm) using gradient elution mode using 0.05% formic acid in water and 0.05% formic acid in acetonitrile as solvents A and B, respectively, 60/40. The run time was 6 min. The mass spectrometer was operated under a positive electrospray ionization condition and a multiple reaction monitoring mode. An excellent linear calibration was achieved in the range of 0.5–1000 ng/mL. An average recovery of belinostat for four quality controls was 72.6% and the recovery of the internal standard at 1000 ng/mL was 67.8%. The intra-day and inter-day precisions for belinostat were ≤8.0 and ≤10.3%, respectively, and their accuracy ranged from 100.2 to 106.7%. No significant matrix effect was identified. In analysis of patient samples, belinostat glucuronide was identified and baseline separated from belinostat. This well-validated assay has been applied for quantification of belinostat in plasma samples within 24 h after the start of infusion for Asian hepatocellular carcinoma patients in a dose escalation study.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , , ,