Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1214098 | Journal of Chromatography B | 2010 | 6 Pages |
A simple, rapid, sensitive and specific ultra performance liquid chromatography tandem mass spectrometry (UPLC–MS/MS) method was developed and validated for the quantification of ethosuximide in human plasma is described. Analyte was chromatographed on a Hypersil Gold C18 column (100 mm × 2.1 mm, i.d., 1.9 μm) with isocratic elution at a flow rate of 0.250 mL/min and pravastatin was used as the internal standard. The assay involves a simple solid-phase extraction procedure of 0.25 mL human plasma and the analysis was performed on a triple-quadrupole tandem mass spectrometer by MRM mode via electrospray ionization (ESI). The method was linear in the concentration range of 0.25–60.0 μg/mL. The lower limit of quantification (LLOQ) was 0.25 μg/mL. The within- and between-day precision and accuracy of the quality control samples were within 10.0%. The recovery was 95.1% and 94.4% for ethosuximide and pravastatin, respectively. The analysis time for each sample was 1.8 min. The method was highly reproducible and gave peaks with excellent chromatography properties.