Article ID Journal Published Year Pages File Type
1214322 Journal of Chromatography B 2010 6 Pages PDF
Abstract

A simple and highly sensitive high-performance liquid chromatography procedure was developed for the determination of carnosine in urine. Carnosine was derivatized with 4-(5,6-dimethoxy-2-phthalimidinyl)-2-methoxyphenylsulfonyl chloride at 70 °C for 15 min in borate buffer (20 mmol l−1, pH 9.0) to produce fluorescent sulfonamides. After hydrolysis of the reaction mixture with formic acid at 100 °C for 15 min, the fluorescent derivative of carnosine was separated on a reversed-phase column with a linear gradient elution using solvents of (A) acetate buffer (0.1 mmol l−1, pH 7.0) and (B) acetonitrile at a flow-rate of 1.0 ml/min and was detected at excitation and emission wavelengths of 318 and 400 nm, respectively. The detection limit of carnosine was 4 fmol at a signal-to-noise ratio of 3. The within-day and day-to-day relative standard deviations were 2.7–4.6% and 0.4–5.2%, respectively. The concentration of carnosine in normal human urine was found to be 4.6–125 nmol (mg creatinine)−1 (mean ± SD: 21.6 ± 26.6 nmol (mg creatinine)−1, n = 20).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , ,