Article ID Journal Published Year Pages File Type
1214783 Journal of Chromatography B 2011 7 Pages PDF
Abstract

A rapid method has been developed to analyse CP 47, 497 in human urine. Urine samples were diluted with water:acetonitrile (90:10, v/v) and sample aliquots were analysed by triple quadrupole tandem mass spectrometry with a runtime of 5 min. Multiple reaction monitoring (MRM) as survey scan was performed. The method was validated in urine, according to an in-house validation protocol based on the criteria defined in Commission Decision 2002/657/EC. Three MRM transitions were monitored. The decision limit (CCα) was 0.01 μg mL−1 and for the detection capability a (CCβ) value of 0.02 μg mL−1 was obtained. The measurement uncertainty of the method was 21%. Fortifying human urine samples (n = 18) in three separate assays, show the accuracy of the method to be between 95 and 96%. The precision of the method, expressed as RSD values for the within-lab reproducibility at the three levels of fortification (0.1, 0.15 and 0.2 μg mL−1) was less than 10% respectively. The method proved to be simple, robust and time efficient. To the best of our knowledge there are no LC–MS methods for the determination of CP 47, 497 with validation data in urine.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,