Abstract
Rapid and precise method for the determination of 8-iso-prostaglandin F2α, an essential marker of the oxidative stress, in exhaled breath condensate (EBC) was developed. The protocol consisted of stable isotope dilution, immunoseparation combined with selective and sensitive LC–ESI-MS/MS operated in multiple reaction monitoring (MRM) mode. The imprecision of the developed method was below 8.8%, the parameter of mean inaccuracy was determined as <9.6% (0–250 pg of 8-iso-prostaglandin F2α/ml EBC). The limit of detection (LOD) was 1 pg/ml EBC and limit of quantification (LOQ) 5 pg/ml EBC. A significant difference in 8-iso-prostaglandin F2α content between the group of asbestosis patients and healthy volunteers was found.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Kamila Syslová, Petr Kačer, Marek Kuzma, Pavlína Klusáčková, Zdena Fenclová, Jindřiška Lebedová, Daniela Pelclová,