Article ID Journal Published Year Pages File Type
1215370 Journal of Chromatography B 2010 9 Pages PDF
Abstract

A total on-line analysis of a target protein from a plasma sample was made using a selective immunoextraction step coupled on-line to an immobilized enzymatic reactor (IMER) for the protein digestion followed by LC–MS/MS analysis. For the development of this device, cytochrome c was chosen as model protein due to its well-known sequence. An immunosorbent (IS) based on the covalent immobilization of anti-cytochrome c antibodies on a solid support was made and an immunoextraction procedure was carefully developed to assess a selective extraction of the target protein from plasma. For the first time, IS was easily coupled on-line with a laboratory-made IMER based on pepsin. The whole on-line device (IS-IMER-LC-MS/MS) allowed the quantification of cytochrome c from 8.5 pmol to 1.7 nmol in buffer medium. Finally, this device was applied to the analysis of only 85 pmol of cytochrome c from plasma with a RSD value lower than 10% (n = 3).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,