Article ID Journal Published Year Pages File Type
1215414 Journal of Chromatography B 2007 6 Pages PDF
Abstract

A rapid, sensitive, and specific LC/MS/MS-based method was developed for determining the concentration of DMXAA in human and mouse plasma. Sample preparation involved a single protein precipitation step using acetonitrile. Separation of DMXAA and 6-isopropoxy-9-oxoxanthene-2-carboxylic acid, the internal standard, was achieved on a Waters X-Terra™ C18 (50 mm × 2.1 mm i.d., 3.5 μm) analytical column using a mobile phase consisting of acetonitrile/10 mM ammonium acetate (55:45, v/v) containing 0.1% formic acid and isocratic flow at 0.2 mL/min for 3 min. The analytes were monitored by tandem mass spectrometry with electrospray positive ionization. Linear calibration curves were generated over the range of 5–3000 ng/mL. The values for precision and accuracy were <9.6%, except at the LLOQ (5 ng/mL) level, which was within 16.8%. Recovery of DMXAA in mouse plasma was >65%. DMXAA was stable through 2 freeze/thaw cycles, to 2 h in mouse plasma or 50% acetonitrile, and on the autosampler to 5.1 h. This method was subsequently used to measure concentrations of DMXAA in mice following intraperitoneal administration.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,