Article ID Journal Published Year Pages File Type
1215478 Journal of Chromatography B 2007 4 Pages PDF
Abstract

The pharmaceutical industry standard for bioanalysis is LC/MS/MS. There are, however, many instances where a single quadrupole detector could successfully be used to provide adequate sensitivity and selectivity for quantitation of drug substances in biological matrices. This paper presents one example of how a single quadrupole detector can be employed in a sensitive and selective analytical method for quantitation of carvedilol. A Synergi Hydro-RP (50 mm × 2 mm i.d.; 4 μm) column was used with acetonitile:water:formic acid mobile phase (32:68:0.01, v/v) at a flow rate of 200 μL/min into a single quadrupole mass spectrometer with an electrospray interface in the positive SIM mode. Using a 300 μL plasma aliquot and a liquid–liquid extraction procedure the limit of quantitation for the assay was 1 ng/mL. The assay utility was demonstrated in the analysis of carvedilol pharmacokinetic profiles in beagle dogs following oral carvedilol administration.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,