Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1215540 | Journal of Chromatography B | 2007 | 7 Pages |
Abstract
An LC-MS/MS assay for the quantitative determination of a new antibacterial agent (AVE6971) has been developed and validated in human white blood cells (WBC). The assay involved a lysing procedure of white blood cells and ultra centrifugation of the extracts. Chromatography was performed on a Supelcosil ABZ+ C18 (2.1 mm Ã 50 mm, 5 μm) column using a mobile phase consisting of methanol/acetonitrile/10 mM ammonium formate mixture (10:30:60, v/v/v) at a flow rate of 0.2 ml/min. The linearity was within the range of 10-10000 ng/ml of extracts, corresponding to 0.5-500 ng of AVE6971 in WBC pellets tubes. The validated lower limit of quantification was 10 ng/ml. The inter- and intra-run coefficients of variation (CV) for the assay were <12.9% and the accuracy were from â9.0 to â1.2%. AVE6971 was stable in WBC for at least 1 month at â75 °C. This assay proved to be suitable for the determination of AVE6971 in WBC from clinical studies.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Arnaud Salvador, Jean-Yves Gautier, Olivier Pasquier, Henri Merdjan,