Article ID Journal Published Year Pages File Type
1215609 Journal of Chromatography B 2007 9 Pages PDF
Abstract

A novel method for the separation and simultaneous determination of urinary d- and l-lactic acid enantiomers by high performance liquid chromatography–tandem mass spectrometry (HPLC/MS/MS) is presented. The chiral separation was optimized on a Chirobiotic teicoplanin aglyocone (TAG) column. Most interestingly, the addition of water in small volume fraction to the polar organic mobile phase was found to significantly improve the chromatography. Calibration curves were linear (r2 > 0.9950) over the range 3–1000 mg/L for l-lactic acid and 0.5–160.8 mg/L for d-lactic acid. The limit of detection (LOD) (S/N = 3) and limit of quantification (LOQ) (S/N = 10) were determined experimentally (n = 3) to be 0.2 and 0.5 mg/L for l-lactic acid and 0.4 and 1.3 mg/L for d-lactic acid, respectively. The normal patient range of l-lactic acid was 1–20 μg/mg creatinine with an elevated value of 85 μg/mg creatinine. For d-lactic acid, the range of normal values were between 0 and 5 μg/mg creatinine with an elevated value of 40 μg/mg creatinine. Finally, the validated method allows for rapid analysis with a total run time of 7.5 min.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,